PROCESS CONTROL & IMPROVEMENT
A calibration certificate is not calibration acceptance
The CMM goes to an accredited external laboratory every year. The report always comes back on time, and the laboratory's label is on the machine. Then the auditor asks: "Who compared the reported error with your acceptance criteria, and where is that record?" The report had simply been filed. Nobody had decided whether the result was acceptable.
What 7.1.5 asks for
7.1.5.1 asks for monitoring and measuring resources that give valid and reliable results, are suitable for the measurement and are maintained, with documented information showing fitness for purpose.
Where measurement traceability is required, 7.1.5.2 asks that equipment be calibrated or verified at specified intervals or before use against standards traceable to international or national standards, be identified so its status is known, and be protected from adjustments, damage or deterioration that would invalidate it.
The aerospace additions are where most findings occur:
- A recall process for equipment requiring calibration or verification.
- A register of monitoring and measuring equipment that includes equipment type, unique identification, location, and the calibration or verification method, frequency and acceptance criteria.
- The scope includes test hardware, test software, automated test equipment and plotters used to produce inspection data, as well as personally owned and customer-supplied equipment used to provide evidence of conformity.
- Calibration or verification carried out under suitable environmental conditions.
- When equipment is found unfit for its intended purpose, a determination of whether previous measurement results were adversely affected, and appropriate action.
The calibration evidence chain
- Register entry
- Recall schedule
- External or internal calibration
- Report review and acceptance decision
- Status identification
- Use
- Impact assessment if the equipment fails
The external report is only step 3. The organization still has to decide whether the reported error is within its own acceptance criteria. If it is not, the next question is what happens to everything measured with that equipment. Laboratory competence is commonly confirmed through ISO/IEC 17025 accreditation and its scope, which many aerospace customers and Nadcap criteria also expect.
A worked example: one worn plug gauge
At fictional ABC Aerospace, an Ø8.00 plug gauge returned from calibration with 0.003 mm wear on the go end. The acceptance criterion was ≤0.002 mm.
| Field | Record |
|---|---|
| Equipment | Ø8.00 plug gauge, GA-0214 |
| Calibration result | Go-end wear 0.003 mm |
| Acceptance criterion | Wear ≤ 0.002 mm |
| Decision | Rejected; removed from use |
| Impact assessment | 18 lots of TH-WB-1001 judged with GA-0214 since the last acceptable calibration |
| Result | A worn go end is undersized and may accept holes slightly below the lower limit. CMM sample data for the period showed a minimum hole diameter of 7.986 mm, above the 7.975 mm limit. Three lots still in-house were re-measured on CMM and all conformed. The 15 shipped lots were judged unaffected; no customer notification needed |
| Decided by | Quality manager |
Not notifying the customer can be the right decision. It is defensible only when the reasoning is recorded, supported by data and made by an authorized person.
A simple impact assessment flow
- Confirm the size and direction of the error and the date of the last acceptable calibration.
- List the part numbers, lots and characteristics judged with that equipment since then.
- Compare the error with the tolerance. If it could accept nonconforming product, mark the lots as affected.
- Re-measure affected product still in-house; evaluate whether shipped product requires customer notification.
- Record the assessment, the result and who decided.
Questions an auditor may ask
- Does the register show the calibration method and acceptance criteria for this equipment? Method is the field most often missing.
- Who accepted this external calibration report, and against which criteria?
- Are operators' personal calipers on the register?
- After the last calibration failure, which product did you trace, and where is the record?
- Is your CMM measurement program treated as measuring equipment? How is it validated?
Certificate ≠ acceptance
A calibration certificate tells you how large the error is. Only the organization can decide whether that error is acceptable for its product. Leaving that judgement to the laboratory means leaving product acceptance to someone else.
About this guide
This article is adapted from a chapter of AS9100D in Practice: From Requirements to Objective Evidence. It refers to AS9100D by clause number and does not reproduce its text. It is not an official interpretation and does not replace the standard, customer requirements or your own procedures. Not affiliated with or endorsed by SAE, IAQG or ISO. Companies, part numbers and events in the example are fictional.
FROM KNOWLEDGE TO PRACTICE
Read the full chapter: AS9100D in Practice includes the full register fields, a recall procedure, two common nonconformity cases and the link to test-software validation.
Check your evidence: AS9100D Internal Audit & Objective Evidence Master Toolkit
AS9100D in Practice is published in Traditional Chinese.
Continue reading
Get email updates from AeroQuality Insight.